baxter sigma spectrum infusion system firmware 6.05 vulnerabilities and exploits
(subscribe to this query)
6.8
CVSSv3
CVE-2014-5431
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. Th...
Baxter Sigma Spectrum Infusion System Firmware 6.05
9.8
CVSSv3
CVE-2014-5432
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue c...
Baxter Sigma Spectrum Infusion System Firmware 6.05
9.8
CVSSv3
CVE-2014-5433
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an malicious ...
Baxter Sigma Spectrum Infusion System Firmware 6.05
9.8
CVSSv3
CVE-2014-5434
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter...
Baxter Sigma Spectrum Infusion System Firmware 6.05