7.8
CVSSv2

CVE-2021-37166

Published: 02/08/2021 Updated: 07/11/2023
CVSS v2 Base Score: 7.8 | Impact Score: 6.9 | Exploitability Score: 10
CVSS v3 Base Score: 7.5 | Impact Score: 3.6 | Exploitability Score: 3.9
VMScore: 694
Vector: AV:N/AC:L/Au:N/C:N/I:N/A:C

Vulnerability Summary

A buffer overflow issue leading to denial of service exists in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.

Vulnerability Trend

Vulnerable Product Search on Vulmon Subscribe to Product

swisslog-healthcare hmi-3 control panel firmware