605
VMScore

CVE-2017-12725

Published: 15/02/2018 Updated: 02/03/2018
CVSS v2 Base Score: 6.8 | Impact Score: 6.4 | Exploitability Score: 8.6
CVSS v3 Base Score: 5.6 | Impact Score: 3.4 | Exploitability Score: 2.2
VMScore: 605
Vector: AV:N/AC:M/Au:N/C:P/I:P/A:P

Vulnerability Summary

A Use of Hard-coded Credentials issue exists in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Vulnerable Product Search on Vulmon Subscribe to Product

smiths-medical medfusion_4000_wireless_syringe_infusion_pump 1.1

smiths-medical medfusion_4000_wireless_syringe_infusion_pump 1.6

smiths-medical medfusion_4000_wireless_syringe_infusion_pump 1.5