Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Vulnerable Product | Search on Vulmon | Subscribe to Product |
---|---|---|
medtronicdiabetes 508_minimed_insulin_pump_firmware - |
||
medtronicdiabetes 522_paradigm_real-time_firmware - |
||
medtronicdiabetes 722_paradigm_real-time_firmware - |
||
medtronicdiabetes 523_paradigm_revel_firmware - |
||
medtronicdiabetes 723_paradigm_revel_firmware - |
||
medtronicdiabetes 523k_paradigm_revel_firmware - |
||
medtronicdiabetes 723k_paradigm_revel_firmware - |
||
medtronicdiabetes 551_minimed_530g_firmware - |
||
medtronicdiabetes 751_minimed_530g_firmware - |