6.9
CVSSv2

CVE-2018-8868

Published: 03/07/2018 Updated: 09/10/2019
CVSS v2 Base Score: 6.9 | Impact Score: 10 | Exploitability Score: 3.4
CVSS v3 Base Score: 6.4 | Impact Score: 5.9 | Exploitability Score: 0.5
VMScore: 614
Vector: AV:L/AC:M/Au:N/C:C/I:C/A:C

Vulnerability Summary

Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Vulnerability Trend

Vulnerable Product Search on Vulmon Subscribe to Product

medtronic 24950_mycarelink_monitor_firmware -

medtronic 24952_mycarelink_monitor_firmware -