6.5
CVSSv3

CVE-2019-18241

Published: 26/11/2019 Updated: 18/12/2019
CVSS v2 Base Score: 3.3 | Impact Score: 2.9 | Exploitability Score: 6.5
CVSS v3 Base Score: 6.5 | Impact Score: 3.6 | Exploitability Score: 2.8
VMScore: 294
Vector: AV:A/AC:L/Au:N/C:P/I:N/A:N

Vulnerability Summary

In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub.

Vulnerability Trend

Vulnerable Product Search on Vulmon Subscribe to Product

philips intellibridge_ec40_firmware

philips intellibridge_ec80_firmware